European Journal of Anaesthesiology



Original Article
(RD) Surgery

Assessment of accelerography with the TOF-GUARD™: a comparison with electromyography


A. A. Dahaba a1c1, P. H. Rehak a1 and W. F. List a1
a1 Department of Anaesthesiology and Department of Biomedicine and Computing, Karl-Franzens University, Graz, Austria

Article author query
dahaba aa   [PubMed][Google Scholar] 
rehak ph   [PubMed][Google Scholar] 
list wf   [PubMed][Google Scholar] 

Abstract

The TOF-GUARD™ is a new device for monitoring the neuromuscular function using acceleration measurement. It is quick and easy to apply and does not require a rigid support for the arm. Forty-one patients were studied to assess the monitoring of vecuronium neuromuscular block (NMB) using accelerography by the TOF-GUARD™ compared with electromyography by the Relaxograph®. Although the mean first twitch (T1%) and mean train-of-four (TOF) ratios measured by the TOF-GUARD™ corresponded to a certain extent with the Relaxograph®, the wide variations of the values for individual patients measured by the TOF-GUARD™ compared with the Relaxograph® and the differences in clinical duration and recovery index between the two monitors do not allow the values of the two monitors to be used interchangeably. The levels at intubation as well as at full recovery of the patients can be assessed equally by the two monitors. Thus, the TOF-GUARD™ is a reliable clinical monitor in daily anaesthesia practice.

(Published Online August 16 2006)
(Accepted February 1997)


Key Words: neuromuscular block; monitoring; neuromuscular function; accelerography; electro-myography; train-of-four; first twitch; equipment.